Oxygen Analysis
Describe the principles of measuring oxygen concentration
As oxygen is a molecule containing two similar atoms, its partial pressure cannot be determined using infrared techniques (unlike CO2). Oxygen content of a gas is instead determined using:
- Paramagnetic analyses
- Fuel Cells
Paramagnetic Analysis
Principles of paramagnetic analysis:
- Oxygen is paramagnetic
This means it is attracted by magnetic fields, but does not propagate the field.- This is because its two unpaired valent electrons have the same spin.
Many other gases weakly repelled by magnetic fields (diamagnetic)
The attraction of a gas mixture to a magnetic field is therefore proportional to its oxygen content
- Many different methods exist which use this property to determine oxygen content
Pressure Method
- Gas tested flows into a tube
- A reference gas flows into a parallel tube
- Both gases then pass through:
- Flow restrictors
- Magnetic field
This is being turned on and off at ~100Hz.
- The gases combine in the magnetic field
- The greater the oxygen content of the gas, the more it will move into the magnetic field
This movement creates a negative pressure behind the gas. - The pressure difference between the tested gas and the reference gas is proportional to the oxygen content of the test gas.
Temperature Method
Used in many modern devices.
- Gas flows through a magnetic field, causing the particles to align
This changes the thermal conductivity of the oxygen molecules. - The change in thermal conductivity of the gas mixture is proportional to the oxygen content
- This is detected by measuring current passing through a heated wire
Pros
- Accurate
- Rapid response time
Modern analysers can identify breath-to-breath variation in FiO2. - Don't require regular calibration
Cons
- Water vapour reduces accuracy
- Interference from other paramagnetic gases
- Nitric oxide
Effect is minimal as nitric is delivered in far smaller volumes than oxygen, and is only weakly paramagnetic.
- Nitric oxide
Fuel Cells
Fuel cells rely on reduction of oxygen to measure oxygen partial pressure. They consist of:
- Oxygen permeable membrane
- KOH solution
This contains:- Lead anode
Lead is consumed as the fuel cell operates. - Gold cathode
- Lead anode
Method
- Oxygen diffuses across the membrane into the potassium hydroxide solution
- At the cathode:
- At the anode:
- The oxygen consumption is proportional to the current generated, which is measured with an ammeter
Pros
- No power required
- Small
- Accurate
Cons
- Will accumulate nitrogen in the presence of N2O
Results in an under-reading of PO2. - Must be replaced after 6-12 months
- Requires regular two-point calibration
21% and 100% oxygen are used. - Relatively slow response time compared to paramagnetic analysers
~20s.
References
- Aston D, Rivers A, Dharmadasa A. Equipment in Anaesthesia and Intensive Care: A complete guide for the FRCA. Scion Publishing Ltd. 2014.