Non-Invasive Blood Pressure
Describe the
invasive andnon-invasive measurement of blood pressureand cardiac outputincluding calibration, sources of errors and limitations
Non-invasive blood pressure measurements is performed with either a:
- Device for Indirect Non-invasive Automatic Mean Arterial Pressure (DINAMAP)
Automatic blood pressure cuff. - Von Recklinghausen's oscillotonometer
"Manual" blood pressure cuff.- Uses two cuffs, and therefore two tubes
DINAMAP
Components:
- One cuff
Performs both arterial occlusion and measurement. - Tubing
- Device for inflating the occlusive cuff and gradually deflating it
- Pressure transducer
- Display
Method:
- Cuff is inflated above SBP
- Cuff deflates at a rate of 2-3mmHg.s-1
When cuff pressure equals: - DBP is calculated from MAP and SBP
Cons
- Requires an appropriately sized cuff
Cuff should be ~20% greater than arm diameter.- Cuffs that are too small will over-read
- Cuffs that are too wide will under-read
- Requires a regular rhythm
- Inaccurate at extremes of blood pressure
- Inaccurate when used more frequently than once per minute
- Inaccurate when the vessel is incompressible
- Heavily calcified vessels
- When applied to forearm/foreleg
- May cause neuropraxia
Von Recklinghausen's Oscillotonometer
Components:
- Two cuffs
- Occlusive cuff
- Measurement cuff
- Tubing
- Device for inflating the occlusive cuff and gradually deflating it
- Aneroid barometer for transducing pressure
- Display
Process:
- Cuff is inflated until the radial pressure is no longer palpable
This is approximates SBP. - Cuff is deflated, and re-inflated to 20mmHg above the estimated SBP
- Cuff is deflated at a rate of 2-3mmHg.s-1 whilst auscultating the brachial artery
When cuff pressure equals:
References
- ANZCA July/August 2000
- Aston D, Rivers A, Dharmadasa A. Equipment in Anaesthesia and Intensive Care: A complete guide for the FRCA. Scion Publishing Ltd. 2014.
- Leslie RA, Johnson EK, Goodwin APL. Dr Podcast Scripts for the Primary FRCA. Cambridge University Press. 2011.